Product Specification SHINYEI KAISHA
Chilled Mirror Dew Point Hygrometer
Shinyei DewStar Series Model. S1-S/M

Shinyei DewStar S1-S/M. Монитор температуры точки росы (температура точки росы, температура, давление, вынос датчика до 300м)

The model S1-S/M Chilled Mirror System is a high performance Chilled Mirror Sensor with two-stage Peltier Thermoelectric mirror cooling. It provides accurate and reliable dew point measurement over an extended range for long periods of time with new ACCU-STAR* mirror maintenance feature.

Shinyei DewStar S1-S/M. Монитор температуры точки росы (температура точки росы, температура, давление, вынос датчика до 300м)

The S1-M is a remote multifunction display monitor.
The S1-M can display two selectable measurements simultaneously, while outputting two selectable analog outputs, one selectable alarm point and any combination of selected measurements on a serial port.
The S1-M also controls the function of the S1-S and timing of the ACCU-STAR feature.
The S1-M can be mounted up to 300m away from the S1-S using low cost cable.
The microprocessor controlled S1-S measures dew point by precisely controlling the temperature of a Rhodium plated mirror at the dew point temperature using a digital control algorithm. A fine layer of dew is grown on the mirror surface and mirror temperature is controlled by a thermoelectric heatpump at the point where the dew layer neither grows nor evaporates, but remains in a stable condition.
A 4 wire 100 Ohm platinum RTD is used to measure the stable dew point temperature.

ACCU-STAR*
A: Automatic
C: Contamination
C: Correction
U: Utility
S Special
T Timed
A Accuracy
R Recovery

An additional port for a temperature probe and pressure sensor allow the S1-S/M to calculate and display
many humidity, temperature and pressure based measurements such as:
Dew Point Deg C or F
Temperature Deg C or F
Pressure kPa, mbar, psia, mmHg
Relative Humidity %RH
Absolute Humidity PPMv, PPMw, g/m3
Specifications:
Dew point range 65deg. C D.P. or less at 25deg. C(below ambient Temperature)
Accuracy +/-0.2 deg. C
Repeatability +/-0.05 deg. C
Hysteresis None
Temperature 100 Ohm PRT 4 wire 0.00385 Ohm/Ohm/deg. C Optional
Inputs Pressure 4 to 20 mA abs. pressure sensor
Analog Dual 0 to 5 VDC and 4 to 20mA (Any parameter)
RS-232 Any parameter Outputs
Alarm 100mA AC/DC Opto-Isolated
Ambient Temp. 0 to 60 deg. C Operating
Conditions Pressure 0 to 1 Mpa
Power 85 to 265 VAC 50/60Hz 100W
Weight 5 kg
S1-S Sensor 360x110x130 mm Size
S1-M Monitor 210x110x30 mm